Model 200 SeriesUsers ManualPublication 5600-002-REV-D
6KDS Model 200 SeriesTo simplify syringe loading, the pusher block can be disengaged from the leadscrew by turn-ing the knob(1) and manually moved alo
7KDS Model 200 SeriesOperating Instructions (Continued)KDS260 PUSH-PULL SYRINGE PUMPTravel directionPusher blockSyringe clampSyringe restAnti-syphon c
8KDS Model 200 SeriesOperating Instructions (Continued)MENU FEATURESWhen the pump is first turned on the LCD will display the last Rate or Volume sett
9KDS Model 200 SeriesOperating Instructions (Continued)SYRINGE DIAMETER ENTRYThe pump must be calibrated by identifying the internal diameter of the s
10KDS Model 200 SeriesOperating Instructions (Continued)UNITS SETTINGprompt: Vol.: 00.00 ml <The units displayed can be changed if required.1) Use
11KDS Model 200 SeriesOperating Instructions (Continued)CHANGE OR REVIEW VOLUME SETTING WHILE RUNNINGWhile the pump continues to run press select to r
12KDS Model 200 SeriesOperating Instructions (Continued)MANUAL STALL SETTING AND MICROLITER SYRINGESA movable collar, located on the rear guide rod of
13KDS Model 200 SeriesOperating Instructions (Continued)CLEARING A STALL CONDITIONShould a stall occur the pump motor is stopped to prevent damage.To
14KDS Model 200 SeriesOperating Instructions (Continued)RS232 COMMANDSRS232 is used for remote computer control of up to 100 pumps, identified with an
15KDS Model 200 SeriesOperating Instructions (Continued)RS232 COMMANDS (continued)Prompts >running in infusion direction< withdrawing: stoppedNA
EU Directives WEEE and RoHSTo Our Valued Customers:We are committed to being a good corporate citizen. As part of that commitment,we strive to maintai
16KDS Model 200 SeriesOperating Instructions (Continued)RS232 COMMANDS (continued)ratei nnnnn u/uSets infusion rate.Where is nnnnn is ., 0 to 9, and
17KDS Model 200 SeriesOperating Instructions (Continued)RS232 COMMANDS (continued)error? Errors 8 - 15 only occur if using pressure switch.8 = overpre
18KDS Model 200 SeriesProgrammable Option– General InformationKEYPAD PROGRAMMABLE PUMPSThe "program mode " is available on all Series 200 mo
19KDS Model 200 SeriesProgrammable Option–FeaturesPARAMETERS WHICH CAN BE PROGRAMMED IN EACH STEPTime durationInfuse or withdrawStart flowrateEnd flow
20KDS Model 200 SeriesKDS Model 200 SeriesProgrammable Option– Menu OperationAfter selecting MODE and then PROGRAM the display will show STEP which le
21KDS Model 200 SeriesProgrammable Option– Menu Operation (Continued)8) Loop (Continued)b) LOOP COUNTAfter setting the initial step number of the loop
22KDS Model 200 SeriesKDS Model 200 SeriesProgrammable Option– Running The ProgramRun The Run key starts the program; the displayed time counts down a
23KDS Model 200 SeriesKDS Model 200 SeriesProgrammable Option– RS232 Commands and ResponsesAll commands and responses in standard pumps remain the sam
KDS Model 200 SeriesKDS Model 200 Series24KDS Model 200 SeriesProgrammable Option– RS232 Commands and Responses(Continued)mode prgmsets pump in progra
Programmable Option– RS232 Commands and Responses(Continued)timeleft?queries time remaining in active stepresponse: xx:xx:xx where hr:min:secnumber? q
KDS Model 200 SeriesContentsSpecifications...
26KDS Model 200 SeriesProgrammable Option– RS232 Commands and Responses(Continued)PROGRAM EXAMPLESyringe selected, 4.70 mm diameterRS232 programmingmo
27KDS Model 200 SeriesRS232 COMMANDSRS232 Format8 data bitsNo parity1 stop (can use 2 stops)Pump uses simple three wire communications - ground, trans
28KDS Model 200 Series549876321TTL SpecificationsPin3 Vss , ground ref.1,6 Controllable output with Programmable models could be used for relay orvalv
29KDS Model 200 SeriesMaintenanceMaintenance is required only for the moving mechanical parts, which should be kept cleanand lubricated. Occasionally,
30KDS Model 200 Series(1) "Air-Tite "All Plastic1 cc 4.70 mm2.5 9.705.0 12.4810 15.8920 20.0030 22.5050 28.90(2) Becton DickinsonInterim, WW
31KDS Model 200 Series(1) "Air-Tite "All Plastic1 cc 4.70 mm2.5 9.705.0 12.4810 15.8920 20.0030 22.5050 28.90(2) Becton DickinsonInterim, WW
32KDS Model 200 SeriesStandard Minimum and Maximum Flow RatesSyringe size Diameter* Minimum Maximum10 µl 0.46 mm 0.001 µl/hr 21.10 µl/min25 µl 0.73 mm
33KDS Model 200 SeriesJapanese Minimum and Maximum Flow Rates(Available in Japanese Models)Syringe size Diameter* Minimum Maximum10 µl 0.46 mm 0.001 µ
34KDS Model 200 SeriesLimited WarrantyKD Scientific Inc. warrants to the first consumer purchaser, for a period of one year from thedate of purchase t
EU Directives WEEE and RoHSTo Our Valued Customers:We are committed to being a good corporate citizen. As part of that commitment,we strive to maintai
KDS Model 200 SeriesFuses...
1KDS Model 200 SeriesSpecificationsModel 200 SeriesSyringe size 10 microliter to 140 milliliterElectrical Rating US model 115 V~, 0.25 ACE model 230 V
2KDS Model 200 SeriesGeneral Safety SummaryPlease read the following safety precautions to ensure proper use of your syringe pump. Toavoid potential h
3KDS Model 200 SeriesGeneral DescriptionThis manual applies to the KDS200 series pumps; infusion pumps, models 200, 220, 250;infusion/withdrawal pumps
4KDS Model 200 SeriesFeatures (continued)4) MODES OF OPERATIONInfusionRate and volume settings: pump infuses to the set volume and stops.Rate setting
5KDS Model 200 SeriesKDS Model 200 SeriesFeatures (continued)7) STALL DETECTIONThe motor is monitored by an optical encoder to confirm the programmed